Pulmonary valve simulation – Fluke Biomedical medSim 300B User Manual
Page 108
medSim 300B
Operators Manual
4-10
The single lumen catheter pullback (aortic valve) simulation begins as follows on
the medSim 300B:
bp2:art
ins
<F1>
<F2>
<F3>
<F4>
bp1:art
eic164.eps
The top line of the display gives the following information:
bp2:art
BP channel 2 arterial blood pressure—120/80
bpl:art
BP channel 1 arterial blood pressure—120/80
2. To simulate insertion of the catheter back into the left ventricle select <F1>-INS.
The medSim 300B returns to the previous display as shown below:
bp2:art
plbk
down
up
<F1>
<F2>
<F3>
<F4>
bp1:lv
###/0
eic165.eps
Note
In the above display "###/0" represents the previously selected left
ventricular pressure value.
3. To exit the medSim 300B aortic valve simulation, press the <up arrow> key until
you view the desired menu.
Pulmonary Valve Simulation
For pulmonary valve simulation, using medSim 300B channels BP3 and BP4/AUX,
connect the medSim 300B to the cardiac catheterization system under test.