Pulmonary valve simulation – Fluke Biomedical medSim 300B User Manual

Page 108

Advertising
background image

medSim 300B

Operators Manual

4-10

The single lumen catheter pullback (aortic valve) simulation begins as follows on

the medSim 300B:

bp2:art

ins

<F1>

<F2>

<F3>

<F4>

bp1:art

eic164.eps

The top line of the display gives the following information:

bp2:art

BP channel 2 arterial blood pressure—120/80

bpl:art

BP channel 1 arterial blood pressure—120/80

2. To simulate insertion of the catheter back into the left ventricle select <F1>-INS.

The medSim 300B returns to the previous display as shown below:

bp2:art

plbk

down

up

<F1>

<F2>

<F3>

<F4>

bp1:lv

###/0

eic165.eps

Note

In the above display "###/0" represents the previously selected left

ventricular pressure value.

3. To exit the medSim 300B aortic valve simulation, press the <up arrow> key until

you view the desired menu.

Pulmonary Valve Simulation

For pulmonary valve simulation, using medSim 300B channels BP3 and BP4/AUX,

connect the medSim 300B to the cardiac catheterization system under test.