Cardiac Output

Cardiac Output Simulation

5

5-5

The purpose of the CI-3 module/cable is:

To facilitate the medSim 300B connection to the cardiac output measurement

device under test.

To simulate the injectate temperature (IT) thermistor at 2 °C.

Of the two connectors on the CI-3 module/cable, the smaller 3-pin connector is for

catheter blood temperature (BT) and is standard on most monitors.

Note

This 3-pin catheter BT connector is compatible with the standard Baxter

(Edwards) BT catheter and similar catheters available from other

manufacturers such as Viggo-SpectraMed and Abbott (Sorenson).

The larger 4-pin connector supplies the simulated injectate temperature.
The 10-turn 100 ke potentiometer enables adjustment of the injectate temperature to

2 °C.
Although this 100 ke range of resistance has proven adequate to simulate the 2 °C

temperature for most brands of cardiac output measurement devices, the 4-pin IT

thermistor connector is not standard on all monitors. Therefore, a general purpose

connector that can be wired to the equipment under test (EUT) injectate cable is also

supplied.

Note

An illustration of the assembly and wiring of the general purpose connector

is located at the end of this chapter. Please note that the injectate

temperature thermistor will have to be cut off the EUT cable to install the

general purpose connector. In addition, an EUT cardiac output cable

modified for testing purposes should not be used for any clinical

application. If the EUT cardiac output cable incorporates an

interconnection (or plug-in) for the injectate temperature bath probe

assembly, modify that plug-in cable only.

Note

To facilitate the interface of the medSim 300B patient simulator to selected

brands of cardiac output measurement devices, optional adapter cables are

offered by Fluke Biomedical.

To evaluate Hewlett Packard Merlin systems, a cardiac output

adapter and a temperature adapter are required.

Gould/SpectraMed Models SP1445 and SP1465 cardiac output

devices each require a specific adapter.

See the current Fluke Biomedical Price List for availability, part number

information and price.

For cardiac output simulation, connect the medSim 300B to the EUT using the supplied

CI-3 module/cable and, if necessary, the general purpose connector.
Set up the EUT for the following simulated parameters:
Catheter Size:

7 F

Injectate Volume:

10 cc

Computational Constant:

0.542