Artifact section, Ecg artifact, Blood pressure/respiration artifact – Fluke Biomedical medSim 300B User Manual

Page 22: Sequences section, Utility functions section, Rs-232 serial port interface

Advertising
background image

medSim 300B

Operators Manual

1-10

Artifact Section

ECG Artifact

Wave Type:

50 Hz, 60 Hz, muscle, or baseline-wander

Leads:

Any wave type can be added to any limb lead, or to the V-leads, or to all leads

Size:

0.25, 0.5, and 1.0 times the Lead II ECG amplitude setting

Blood Pressure/Respiration Artifact

Respiration artifact can be injected into any blood pressure waveform. Arterial and left

ventricle waveforms are modulated by the selected respiration rate at 5 % or 10 %. All

others have respiration added to them at 5 mmHg or 10 mmHg.

Sequences Section

Sequences step through a series of preprogrammed selections.
There are two types of sequences—autosequences and step-sequences. Autosequences

run automatically with no display interaction; step-sequences can be started and stopped

by the operator. There are three preprogrammed step-sequences and four

user-programmable step sequences.

Note

See the "Operation" chapter for more information and a list of the

sequences.

Utility Functions Section

View Angle:

Down, up and save; saved permanently

Auto Power Off Time:

0.5, 1.0, 2.0, 4.0 and 8.0 hours; and off.

This is reset each time a key is pressed; saved permanently

Key Beep:

On and off; not saved

RS-232 Baud Rate:

300, 600, 1200, and 2400 baud; saved permanently

Save:

Saves all current settings as power-on conditions

RS-232 Serial Port Interface

The medSim 300B can be interfaced to a personal computer or to the Fluke Biomedical

medTester 5000B. Most functions can be controlled remotely except for slide switch

controlled functions.

Note

A special cable is required and is available from Fluke Biomedical. Refer

to the current Fluke Biomedical Price List for availability, part number

information and price.

Baud rates are 300, 600, 1200, and 2400.