Option 1

Intra-Aortic Balloon Pump (IABP) Augmentation

4

4-33

Late Deflation

Inflation

Deflation

Dicrotic Notch

Assist Ratio = 1:2

15% Decrease

Assisted

Waveform

Unassisted

Waveform

eic209.eps

Figure 4-29. medSim 300B Simulated Late Balloon Deflation

Inflation begins at the dicrotic notch (aortic valve closure).

Late deflation causes a 15 % decrease in the following systolic peak pressure.

Normal Sinus Rhythm Simulation

For simulated normal sinus rhythm (NSR) waveforms:

1. Select <F3>-BP from the FUNCTION SELECT MENU for the following:

BP CHANNEL SELECT

1

2

3

>

<F1>

<F2>

<F3>

<F4>

4

eic210.eps

2. Press <F4>-4:

BP 4

WAVE

STAT

IABP

<F1>

<F2>

<F3>

<F4>

eic211.eps

3. Press <F3>-IABP:

INTRA AORTIC BALLOON PUMP

NSR

ARRH

PACE

<F1>

<F2>

<F3>

<F4>

eic212.eps