Fluke Biomedical medSim 300B User Manual

Page 75

Advertising
background image

Operation

Operating the medSim 300B

3

3-47

2. From this menu, press the function key that corresponds to the user sequence that

is to be programmed.
When a sequence choice is made, the display returns to the FUNCTION

SELECT MENU and an asterisk (*) at each end of the second line indicates that

the "make" sequence mode is running. The menu display looks like this:

FUNCTION SELECT

ECG

RESP

BP

TEMP

<F1>

<F2>

<F3>

<F4>

*

*

>

eic112.eps

The following example shows how a waveform segment of a user step-sequence is saved

and placed into a sequence. As an illustration, an ECG baseline BPM setting with its run

time is placed into a user step-sequence.
After a sequence (<F1>-SEQ1, <F2>-SEQ2, <F3>-SEQ3, or <F4>-SEQ4) has been

chosen from the SEQUENCE TO MAKE MENU:

1. Press <F1>-ECG from the FUNCTION SELECT MENU:

ECG

BASE

PERF

ARRH

PVC

<F1>

<F2>

<F3>

<F4>

*

*

>

eic113.eps

2. Press <F1>-BASE:

BASELINE ECG

BPM

AMP

ST

AXIS

<F1>

<F2>

<F3>

<F4>

*

*

>

eic114.eps

3. Press <F1>-BPM:

baseline bpm

30

60

80

120

<F1>

<F2>

<F3>

<F4>

*

*

>

eic115.eps

Advertising