Cardiac output, Temperaturešvariable and fixed, Blood pressure – Fluke Biomedical medSim 300B User Manual

Page 34: Respiration, Rear panel, Battery eliminator, Operating the medsim 300b

Advertising
background image

medSim 300B

Operators Manual

3-6

V PACE

This jack is not active.

SYNC/A PACE

The jack for input of a pulse from a defibrillator for the interactive defibrillator training

mode, and for use as the Intra-Aortic Balloon Pump (IABP) signal input for arterial

waveform augmentation.

Bottom Panel

Battery Compartment

Houses two 9-volt alkaline batteries. For replacement information, refer to instructions

contained in the POWER section at the beginning of this chapter.

ROMS and CAL Switch Access

The calibration-enable switch and the software ROMs are housed behind this panel.

Operating the medSim 300B

Initialization

Turn on the instrument by pressing the "ON/OFF" key, and a momentary display of the

firmware version appears, as shown below:

MEDSIM 300 B 3.00

FLUKE BIOMEDICAL INC

<F1>

<F2>

<F3>

<F4>

eic002.eps

Note

If options are installed, they are indicated on the top line of the display. The

Cardiac Output Option is indicated by "CO" and Option 1 is indicated by

"OP1".

If "power off time" has been manually set from the UTILITY MENU, a momentary

display appears indicating time in hours as shown below:

POWER OFF TIME IS ## HR

<F1>

<F2>

<F3>

<F4>

eic003.eps

Note

To set "power off time" refer to the Function Select - Utility Menu section

later in this chapter.

Advertising