Fluke Biomedical medSim 300B User Manual
Page 58
medSim 300B
Operators Manual
3-30
2. Press the corresponding function key for the desired BP 1 dynamic waveform.
Waveform selections and their descriptions follow.
Selections Descriptions
atm
Atmosphere - 0 mmHg.
art
Arterial at 120/80 mmHg.
lv
Left ventricle at 120/0 mmHg.
cvp
Central venous pressure at 15/10 mmHg.
rv
Right ventricle at 25/0 mmHg.
pa
Pulmonary artery at 25/10 mmHg.
paw
Pulmonary artery wedge at 10/2 mmHg.
To select BP 1 static pressures, press <F2>-STAT from the BP 1 MENU for the
following display:
bp 1 static:
## mmHg
down
up
exc
auto
<F1>
<F2>
<F3>
<F4>
eic073.eps005A
Follow the instructions for the step menu functions presented earlier in this chapter.
BP 2 dynamic waveforms and static pressures are identical to BP 1. To select BP 2, press
<F2>-2 from the BP CHANNEL SELECT MENU to access the following menu:
BP 2
WAVE
STAT
<F1>
<F2>
<F3>
<F4>
eic074.eps
To select a BP 2 dynamic waveform:
1. Press <F1>-WAVE to access the two menu pages as shown:
bp 2 waveforms
atm
art
lv
cvp
<F1>
<F2>
<F3>
<F4>
>
eic075.eps