Equipment required for calibration, Calibration setup – Fluke Biomedical medSim 300B User Manual

Page 154

Advertising
background image

medSim 300B

Operators Manual

6-6

Equipment Required for Calibration

To begin the "on-site" calibration process, make sure you have the following equipment:

Digital multimeter- Keithley 196

*

Measurement Type

Measurement

Range

Accuracy (<±) Meter Output

Current

VDC

0-6 mV

10 µV

N/A

VDC

0-12 mV

20 µV

N/A

VDC

0-120 mV

200 µV

N/A

VDC

10 V

10 mV

N/A

Resistance 500

e

0.1

e

N/A

Resistance 15-25

ke 50

e

1-10

µV

Resistance 1-5

ke 5

e

10-100

µV

DC power supply - Instek Model PS-3225D

*

: 10 V ±10 mV @ 30 mA

Calibration cables

**

Part No.

Blood Pressure

2199081

Temperature

2199278

Cardiac Output

2199269

Note

*

Equivalent equipment is required to meet or exceed listed specifications.

Note

**

Cables are available from Fluke Biomedical. Refer to the current Fluke

Biomedical Price List for availability, part number information and price

Calibration Setup

The medSim 300B has no internal adjustments. All calibration adjustments are completed

using the medSim 300B keyboard. Calibration constants are stored in the nonvolatile

memory (EEPROM) within the medSim 300B.

1. Remove the Quality Seal from the bottom panel.
2. Then remove the three screws holding the metal plate and remove the metal

plate.

3. Locate the "CAL" switch near the rear of the opening.
4. Set the top panel switches as follow:

BASELINE IMPEDANCE

500

LEAD SELECT

II

SENSITIVITY 40

VAR TEMP-PROBE

700

CARDIAC OUTPUT/VAR

TEMP

VAR TEMP