Mitral valve simulation – Fluke Biomedical medSim 300B User Manual

Page 114

Advertising
background image

medSim 300B

Operators Manual

4-16

120

60

mmHg

10

LV = 120/0

LVEDP = 10 mmHg

PAW = 10/2

eic175.eps

Figure 4-11. Simulated Waveforms for Normal Blood Pressure Relationships between the Left Atrial

Pressure and the Left Ventricular End Diastolic Pressure

120

mmHg

10

LV = 120/0

LVEDP = 10 mmHg

PAW = 36/24

PAW = 34/22

PAW = 31/21

PAW = 29/19

PAW = 26/18

PAW = 10/2

eic176.eps

Figure 4-12. Simulated Systolic Pressure Gradients Indicating a Defective Mitral Valve

Note

Increases in left atrial pressure (represented by PAW pressure) over left

ventricular end diastolic pressure (LVEDP) simulate a defective mitral

valve.

Mitral Valve Simulation

For mitral valve simulation, using medSim 300B channels BP1 and BP3, connect the

medSim 300B to the cardiac catheterization system under test.