Performance Checks/Calibration

Calibration

6

6-7

5. Plug the medSim 300B into the battery eliminator. Press and hold down the

"CAL" switch.

6. Turn on the medSim 300B and continue to hold down the "CAL" switch until the

display reads:

TEST FUNCTION SELECT

HDWR

RCAL

CAL

EXIT

<F1>

<F2>

<F3>

<F4>

eic257.eps

7. Press <F3>-CAL. The display shows the first calibration step.

Calibration Procedure

During calibration, the top line of the display shows the name of the calibration step,

where to connect the digital multimeter (DMM) and the value to adjust for.
Some of the steps require only that a DMM reading be taken before proceeding to the

next step.
For steps that require adjustment, the menu selections are displayed on four menu pages,

which are accessed using the <right arrow> and <left arrow> keys. Press the function key

that relates to the desired selection.
To decrement and increment, use the <F1> and <F2> keys. The <up arrow> key saves

the setting and jumps back a step.
Menu selections and their functions follow:

Selections Functions
sdn slow

decrement

sup slow

increment

mdn medium

decrement

mup medium

increment

fdn fast

decrement

fup fast

increment

next

moves to the next step

skip

moves to the next step without saving the new setting

esc

exits the calibration procedure without saving the new setting

save

exits the calibration procedure and saves the new setting

1. For the "ECG: REF" step, record the reading (including polarity).
2. Then on the next step—"ECG: 5 mV"—adjust the unit for a reading of 5 mV

plus the reference step reading. This can also be done by zeroing the DMM on

the reference step.

3. For "PACE: REF", record the reading as above. Then for the "ATRIAL PACE:

REF V" and "VENT PACE: REF V", adjust the unit to the "PACE REF" voltage.