medSim 300B

Operators Manual

3-22

Function Select - ECG - Pacemaker Menu

To access the PACEMAKER MENU:

1. Press <F1>-PACE from page 2 of the ECG MENU.

The three main pacemaker parameter selections are pace wave, atrial pacer, and

ventricle pacer, and are displayed as shown:

PACEMAKER

ATR

WAVE

VENT

<F1>

<F2>

<F3>

<F4>

eic046.eps

2. Press <F1>-WAVE from the PACEMAKER MENU to access the PACE WAVE

MENU:

pace wave

asnc

nsr

dem1

dem2

<F1>

<F2>

<F3>

<F4>

>

eic047.eps

pace wave

nonc

avsq

nonf

<F1>

<F2>

<F3>

<F4>

<

eic048.eps

In the PACE WAVE MENU, the atrial-ventricular sequential waveform uses

both atrial and ventricular pacemakers. All others use the ventricular pacemaker.

The pacemaker rate is 75 BPM.

Note

The ventricular pacemaker triggers a wide QRST wave similar to LBBB

since the ventricular pacemaker is in the right ventricle.

Descriptions of the pacemaker waveform selections follow:

Selections Descriptions
nsr

Normal sinus rhythm. Provides a convenient way to return to normal

sinus rhythm.

asnc

Asynchronous. At 75 BPM, continuously paced.

dem1

Demand 1. Mostly paced. A series of 40 paced beats and 20 normal

beats. Pacemaker kicks in when normal beat is 20 % late and kicks

out when normal beat comes in time for an 80 BPM rate.