Fluke Biomedical medSim 300B User Manual
Page 63
Operation
Operating the medSim 300B
3
3-35
8. To select BP 4, press <F4>-4 from the BP CHANNEL SELECT MENU to
access the following menu:
BP 4
WAVE
STAT
<F1>
<F2>
<F3>
<F4>
eic089.eps
Note
Illustrations may not match your medSim 300B display if additions, options,
or expansion modules have been installed.
To select a dynamic waveform from BP 4:
1. Press <F1>-WAVE to access the two menu pages as shown:
bp 4 waveforms
atm
art
lv
rv
<F1>
<F2>
<F3>
<F4>
>
eic090.eps
bp 4 waveforms
pa
paw
t30
t300
<F1>
<F2>
<F3>
<F4>
<
eic091.eps
2. Press the corresponding function key to select a BP 4 dynamic waveform.
Waveform selections and their descriptions follow:
Selections Descriptions
atm
Atmosphere - 0 mmHg.
art
Arterial at 120/80 mmHg.
lv
Left ventricle at 120/0 mmHg.
rv
Right ventricle at 25/0 mmHg.
pa
Pulmonary artery at 25/10 mmHg.
paw
Pulmonary artery wedge at 10/2 mmHg.
t30
Triangle, 30 mmHg, 2 Hz.
t300
Triangle, 300 mmHg, 2 Hz.