Operation

Operating the medSim 300B

3

3-35

8. To select BP 4, press <F4>-4 from the BP CHANNEL SELECT MENU to

access the following menu:

BP 4

WAVE

STAT

<F1>

<F2>

<F3>

<F4>

eic089.eps

Note

Illustrations may not match your medSim 300B display if additions, options,

or expansion modules have been installed.

To select a dynamic waveform from BP 4:

1. Press <F1>-WAVE to access the two menu pages as shown:

bp 4 waveforms

atm

art

lv

rv

<F1>

<F2>

<F3>

<F4>

>

eic090.eps

bp 4 waveforms

pa

paw

t30

t300

<F1>

<F2>

<F3>

<F4>

<

eic091.eps

2. Press the corresponding function key to select a BP 4 dynamic waveform.

Waveform selections and their descriptions follow:

Selections Descriptions
atm

Atmosphere - 0 mmHg.

art

Arterial at 120/80 mmHg.

lv

Left ventricle at 120/0 mmHg.

rv

Right ventricle at 25/0 mmHg.

pa

Pulmonary artery at 25/10 mmHg.

paw

Pulmonary artery wedge at 10/2 mmHg.

t30

Triangle, 30 mmHg, 2 Hz.

t300

Triangle, 300 mmHg, 2 Hz.