Cardiac Output

Overview

5

5-3

Overview

The Cardiac Output Option electronically simulates the dynamic temperature changes in

the patient's blood during thermal dilution cardiac output measurement.
Thermal dilution cardiac output measurements are based upon the heat exchange between

the patient's blood and a known volume of chilled saline injected into the heart. Cardiac

output is expressed in liters per minute (L/min) and ranges between 3 L/min and 7 L/min

for normal adults.
Current cardiac output measurement devices have additional technical capabilities to

ensure that the most accurate readings are attained. These capabilities include the ability

to average a series of readings to eliminate variations because of artifacts, as well as

reject readings because of clinician technique or the patient's underlying cardiovascular

disease.

Note

Cardiac output measurement devices using different techniques (such as

Fick dye injection, Doppler ultrasonography and bioimpedance) are

neither addressed nor intended for use with the Fluke Biomedical

medSim 300B Patient Simulator.

Specifications

The Fluke Biomedical medSim 300B Cardiac Output Option is designed to test the

features and specifications of the cardiac output patient monitors identified above.
The cardiac output simulation is based upon the electrical characteristics of the American

Edwards Catheter 93A-131-7 F which has a 7 F-catheter diameter and uses an injectate

volume of ten cubic centimeters (10 cc). Accordingly, cardiac output values are

programmed at fixed readings of 3 L/min, 5 L/min and 7 L/min; 3 defective readings; and

3 series of readings with variations.

Note

While the Cardiac Output Option is based upon the American Edwards

Catheter, it is compatible with comparable adult cardiac output

measurement catheters, such as Gould/SpectraMed and Abbott (Sorenson).

Specifications for the Cardiac Output Option follow:
Simulated Catheter Type:

American Edwards Catheter 93A-131-7 F

Computational Constant:

0.542

Catheter Size:

7 F

Injectate Volume:

10 cc

Baseline Temperature:

36 °C, 37 °C, and 38 °C

Injectate Temperature:

2 °C

Waveforms:

3 L/min, 5 L/min, and 7 L/min

Interrupted

injectate

Left ventricular shunt

Slow

injectate