medSim 300B

Operators Manual

1-4

medSim 300B Instrument Specifications

ECG Section

The ECG configuration is 12-lead with independent outputs for each signal lead

referenced to RL, and provides:

A "High Level ECG" output.

Selectable parameters including rate, amplitude, ST segment deviation, axis

deviation, neonatal waveforms, performance waveforms, R-wave detection, PVC

focus, PVC timing, PVCs per minute, arrhythmias (32), pacemaker amplitude,

and pacemaker width.

Input for interactive defibrillator training and simulation.

Normal Sinus Rhythm ECG Waveform Selections

Rate:

30, 60, 80, 120, 160, 200, 240, and 300 BPM

Accuracy:

1

%

Amplitude: Lead

II

0.05 mV to 0.5 mV in 0.05 mV steps

0.50 mV to 5.5 mV in 0.25 mV steps

The high level output is approximately 0.2 V per mV of

the low level output. Selection is either manual or

automatic at stepped intervals of 6 seconds

Accuracy: 5 % on Lead II into a 100 Hz low pass filter

Impedance to the RL Lead:

Limb leads are selectable at 500 e, 1000 e, 1500 e and

2000 e. The V-leads are 1000 e

Accuracy:

5%

ST Segment Deviation

(Elevation and Depression): Lead II at 1 mV

+0.8 mV to -0.8 mV in 0.1 mV steps, also +0.05 mV and
-0.05 mV

Both manual and auto-stepping modes. Auto-step time

interval is 12 seconds

Operates at ECG rates from 30 to 160 BPM

Axis Deviation:

Horizontal, vertical and intermediate heart positions