Menu overview, Main menu, Movement through menus – Fluke Biomedical medSim 300B User Manual

Page 36: Executable parameter values, Step menus

Advertising
background image

medSim 300B

Operators Manual

3-8

In the step menu, the top display line indicates the name of the parameter being

modified and a value for it. If the value shown (illustrated above as "#.##") is

active, it flashes.

2. Press <F1>-DOWN or <F2>-UP to decrement or increment the displayed value.
3. Press <F3>-EXC to execute the displayed value.
4. Press <F4>-AUTO and the bottom display line changes as follows:

ecg amp, lead II: #.## mv

down

up

strt

man

<F1>

<F2>

<F3>

<F4>

eic005.eps

In this mode, use of the <Fl>-DOWN key and the <F2>-UP key decrements and

increments the displayed value, and immediately executes the selected value

without requiring a separate "execute" key.

Note

Use of the <F4>-MAN key returns to the previous mode.

5. To activate the automatic stepping mode, press the <F3>-STRT key and the

display changes to:

ecg amp, lead II: #.## mv

down

up

stop

<F1>

<F2>

<F3>

<F4>

eic006.eps

At this point, the parameter is automatically stepping at a predetermined rate

(refer to the "Specifications" chapter). Upon reaching either limit, the stepping

switches directions and continues.

6. Choose <F1>-DOWN or <F2>-UP to reverse the incrementing and decrementing

of the values.

7. Press <F3>-STOP to stop the stepping and return to the manual mode.
8. Pressing the <up arrow> also stops automatic stepping and moves you back up

through the menu.
Only one automatic stepping mode can be active at a time.