Temperature section, Cardiac output section – Fluke Biomedical medSim 300B User Manual

Page 21

Advertising
background image

Chapter Title

medSim 300B Instrument Specifications

1

1-9

Channel 3 (mmHg):

Atmosphere (0)

Central venous pressure = 15/10

Right ventricle = 25/0

Pulmonary artery = 25/10

Pulmonary artery wedge = 10/2

Static: -10, -5, 0, 5, 10, 20 and 30

Manual or auto-stepping at 12 second intervals

Swan-Ganz: start, insert, inflate, deflate and remove

Channel 4 (mmHg):

Atmosphere (0)

Arterial = 120/80

Left ventricle = 120/0

Right ventricle = 25/0

Pulmonary artery = 25/10

Pulmonary artery wedge = 10/2

Triangle = 30, 2 Hz

Triangle = 300, 2 Hz

Static: -10, -5, 0, 20, 40, 80, 100, 200, 250 and 300

Manual or auto-stepping at 12 second intervals

Temperature Section

Variable Temperature:

34 °C, 37 °C, 40 °C, hyperthermia, hypothermia, and spike

Note

See the "Operation" chapter for descriptions of waveforms.

Note

"VAR TEMP" is inoperative during the Cardiac Output simulation.

Fixed Temperature:

Fixed at 37 °C

Channel 2's accuracy is 0.1 °C

Probe Compatibility:

400 and 700 series YSI types

Isolation:

Variable temperature is electrically isolated from the rest of the

instrument except Cardiac Output. Fixed temperature is isolated from

all other outputs.

Cardiac Output Section

See the "Cardiac Output" option chapter in this manual.