Intra-aortic balloon timing, Normal inflation, Early inflation – Fluke Biomedical medSim 300B User Manual

Page 129

Advertising
background image

Option 1

Intra-Aortic Balloon Pump (IABP) Augmentation

4

4-31

Intra-Aortic Balloon Timing

Characteristics—normal inflation, early inflation, late inflation, early deflation and late

deflation—of the medSim 300B balloon timing simulations are illustrated below.
These examples show the medSim 300B responses to an IABP assist ratio of 1:2 in which

every second pressure waveform is augmented.

Normal Inflation

Dicrotic Notch

Inflation

Deflation

Assist Ratio = 1:2

Unassisted
Waveform

Assisted
Waveform

eic205.eps

Figure 4-25. medSim 300B Simulated Normal Waveform with the IABP Properly Set

Inflation begins at the dicrotic notch (the small increase on the downslope of the

arterial waveform that is caused by the closure of the aortic valve).

Deflation ends just before the next arterial pressure waveform.

Early Inflation

Dicrotic Notch

Inflation

Deflation

Assist Ratio = 1:2

Unassisted
Waveform

Assisted
Waveform

eic206.eps

Figure 4-26. medSim 3008 Simulated Early Balloon Inflation

Inflation begins prior to the dicrotic notch (aortic valve closure).

Deflation ends just before the next arterial pressure waveform.

Note

The dicrotic notch is not visible because balloon inflation occurs prior to

the aortic valve closure.

Advertising