Fluke Biomedical medSim 300B User Manual

Page 83

Advertising
background image

Operation

Operating the medSim 300B

3

3-55

At this point, defibrillation can be accomplished either by a defibrillator signal

input or by using the CARDIOVERSION MENU.

3. When defibrillating by a defibrillator pulse signal input through the "SYNC/A

PACE" jack, the timing relationship to the ECG R-wave peak is checked.
If the pulse comes within ±100 ms, defibrillation is successful and the ECG

converts to normal sinus rhythm. The display then shows the following:

cardioversion:

sync okay

<F1>

<F2>

<F3>

<F4>

eic134.eps

and, after 3 seconds, changes to:

cardioversion:

converted

start

hold

<F1>

<F2>

<F3>

<F4>

eic135.eps

This is the end of the sequence. It can be restarted by pressing <Fl>-START.

4. If the pulse signal is outside ±100 ms, defibrillation is late and the ECG changes

to ventricular fibrillation. The display shows:

cardioversion:

late

hold

<F1>

<F2>

<F3>

<F4>

eic136.eps

then changes to:

cardioversion:

vfib

hold

defib

start

<F1>

<F2>

<F3>

<F4>

eic137.eps

5. Press <F3>-DEFIB or input a pulse signal through the rear panel "SYNC/A

PACE" jack to defibrillate the simulator and convert the ECG to normal sinus

rhythm.

Advertising