Fluke Biomedical medSim 300B User Manual
Page 121
Advertising
Option 1
Fetal/Maternal ECG with Intrauterine Pressure
4
4-23
Note
Maternal R-wave is reduced 50 % when fetal heart rate (ECG) is added.
Fetal heart rate = 140 BPM; maternal heart rate = 80 BPM.
For intrapartum fetal and maternal heart rate simulation, connect the medSim 300B to the
fetal monitor under test.
Follow the diagram below:
Intrauterine Blood
Pressure Input
Maternal Thigh
Reference Plate
Fetal Scalp Electrode
Jumper Wires
Fetal
Monitor
medSim 300B
Fetal ECG Input
BP Cable
BP4/AUX Output
ON WT BK
eic187.eps
Figure 4-18. medSim 3008 Connections for Fetal Heart Rate Simulation
Turn both units on.