medSim 300B

Operators Manual

4-6

3. To zero all four of the medSim 300B blood pressure channels, press <F3>-BP

from page 1 of the FUNCTION SELECT MENU, for the following two pages of

display:

BP CHANNEL SELECT

2

3

4

>

<F1>

<F2>

<F3>

<F4>

1

eic158.eps

BP CHANNEL SELECT

<

zero

CATH

<F1>

<F2>

<F3>

<F4>

eic159.eps

4. Press <F1>-ZERO from the second page of the menu.

This sets all the medSim 300B blood pressure channels to zero.

Note

Necessary to complete only once after connecting the medSim 300B to the

equipment under test (EUT).

5. Now zero the blood pressure channels to be used on the EUT.

To begin aortic valve simulation:

1. Choose <F2>-CATH from page 2 of the BP CHANNEL SELECT MENU.

The medSim 300B display shows:

CATHETERIZATION

aort

mitr

pulm

<F1>

<F2>

<F3>

<F4>

eic160.eps