Fluke Biomedical medSim 300B User Manual
Page 104
medSim 300B
Operators Manual
4-6
3. To zero all four of the medSim 300B blood pressure channels, press <F3>-BP
from page 1 of the FUNCTION SELECT MENU, for the following two pages of
display:
BP CHANNEL SELECT
2
3
4
>
<F1>
<F2>
<F3>
<F4>
1
eic158.eps
BP CHANNEL SELECT
<
zero
CATH
<F1>
<F2>
<F3>
<F4>
eic159.eps
4. Press <F1>-ZERO from the second page of the menu.
This sets all the medSim 300B blood pressure channels to zero.
Note
Necessary to complete only once after connecting the medSim 300B to the
equipment under test (EUT).
5. Now zero the blood pressure channels to be used on the EUT.
To begin aortic valve simulation:
1. Choose <F2>-CATH from page 2 of the BP CHANNEL SELECT MENU.
The medSim 300B display shows:
CATHETERIZATION
aort
mitr
pulm
<F1>
<F2>
<F3>
<F4>
eic160.eps