Single and dual lumen catheter simulation – Fluke Biomedical medSim 300B User Manual

Page 105

Advertising
background image

Option 1

Cardiac Catheterization

4

4-7

2. Press <F1>-AORT for the following menu display:

bp2:art

plbk

down

up

<F1>

<F2>

<F3>

<F4>

bp1:lv

120/0

eic161.eps

The top line of the display gives the following information:

bp2:art

BP channel 2 arterial pressure-120/80

bp1:lv

BP channel 1 left ventricular pressure

120/0

Left ventricular pressure setting (variable)

The variable systolic blood pressure selections for left ventricular pressure are:

120/0

Normal (medSim 300B default value)

126/0

5 % above arterial blood pressure

132/0

10 % above arterial blood pressure

138/0

15 % above arterial blood pressure

144/0

20 % above arterial blood pressure

180/0

50 % above arterial blood pressure

3. Press <F4>-UP or <F3>-DOWN on the medSim 300B to increase or decrease the

systolic blood pressure value for the left ventricle on blood pressure channel 1

(bp 1:lv).

Single and Dual Lumen Catheter Simulation

The medSim 300B simulates blood pressure waveforms obtained with single and dual

lumen catheters.
When the medSim 300B begins aortic simulation, it simultaneously represents two types

of measurements:

A single lumen catheter with the lumen placed in the left ventricle.

A dual lumen catheter with the distal lumen placed in the left ventricle and the

proximal lumen in the aorta (one lumen on either side of the aortic valve).