RS-232 Serial Port Operation

3

3-61

AXHRZ Horizontal

axis

AXVER Vertical

axis

Neonatal:

NEOOFF Neonatal,

off

NEOON Neonatal,

on

ECG Performance Waveforms:

ZERO Zero

output

PUL

Pulse at 4 s, 1 mV

SQU

Square waveform, 2 Hz at 1 mV.

TRI

Triangle waveform, 2 Hz at 1 mV

SINn Sine

waveform

n = 0.05, 0.5, 1, 10, 25, 30, 40, 50, 60 and 100

R-wave Width:

RWn

Width = n

n = 8, 12, 20, 40, 60, 80, 100, 120, 140, 160, 180 and 200

Supraventricular Arrhythmias:

AF1

Atrial fibrillation 1

AF2

Atrial fibrillation 2

AFL Atrial

flutter

SINA Sinus

arrhythmia

MB80

Missing beat at 80 BPM

MB120

Missing beat at 120 BPM

PAT

Paroxysmal atrial tachycardia.

NOD Nodal

rhythm

SVT Supraventricular

tachycardia

Premature Arrhythmias:

PAC Premature

atrial

contraction

PNC

Premature nodal contraction

PVC

Premature ventricular contraction

MF1

Multifocal PVCs 1

MF2

Multifocal PVCs 2

MF3

Multifocal PVCs 3

Ventricular Arrhythmias:

BIG Bigeminy

TRG Trigeminy

PAIR

Pair of PVCs