medSim 300B

Operators Manual

4-12

Now zero the blood pressure channels to be used on the EUT.

To begin pulmonary valve simulation:

1. Choose <F2>-CATH from page 2 of the BP CHANNEL SELECT MENU.

The medSim 300B display shows:

CATHETERIZATION

aort

mitr

pulm

<F1>

<F2>

<F3>

<F4>

eic169.eps

2. Press <F3>-PULM for the following menu display:

bp3:pa

plbk

down

up

<F1>

<F2>

<F3>

<F4>

bp4:rv

25/0

eic170.eps

The top line of the display gives the following information:

bp3:pa BP channel 3 pulmonary arterial pressure—25/10
bp4:rv BP channel 4 right ventricular pressure
25/0

Right ventricular pressure setting (variable)

The variable systolic blood pressure selections for right ventricular pressure are:

25/0

Normal (medSim 300B default value)

26/0

5 % above arterial blood pressure

27.5/0 10 % above arterial blood pressure
29/0

15 % above arterial blood pressure

30/0

20 % above arterial blood pressure

37.5/0 50 % above arterial blood pressure

3. Press <F4>-UP or <F3>-DOWN on the medSim 300B to increase or decrease the

systolic blood pressure value for the right ventricle on blood pressure channel 4

(bp4:rv).

Single and Dual Lumen Catheter Simulation

The medSim 300B simulates blood pressure waveforms obtained with single and dual

lumen catheters.
When the medSim 300B begins pulmonary simulation, it simultaneously represents two

types of measurements:

A single lumen catheter with the lumen placed in the right ventricle.