Power, Power-on conditions, Panel descriptions, controls and connectors – Fluke Biomedical medSim 300B User Manual

Page 32

Advertising
background image

medSim 300B

Operators Manual

3-4

Parameter Specification
Artifact All

Off

Power-off time

Off

Key beep

On

Baud rate

2400

Cardiac output baseline

37 °C

Note

These power-on conditions can be changed by utilizing the "save" function

in the UTILITY MENU.

Panel Descriptions, Controls and Connectors

Top Panel

LCD Display and Keyboard

The LCD display-2 lines by 24 characters—displays messages, and provides information

necessary for setting output levels and selecting waveform types.
The keyboard has the following eight pressure-sensitive keys: four function keys (F1, F2,

F3, and F4), three arrow keys (left, up and right), and the "ON/OFF" key.

ECG Leads

Located above the LCD display are ten ECG lead-binding posts that are compatible with

disposable snaps, 3.2 mm or 4 mm electrodes, and banana plugs. All leads have

independently generated signals.

RS232

For remote control of the patient simulator, the RS232 serial port interface is used to

connect the medSim Controller Option to a computer or terminal. For information

regarding the operation of the RS-232 interface, see the RS-232 SERIAL PORT

OPERATION section later in this chapter.

High Level ECG

The "HIGH LEVEL ECG" port is a 1/4" standard phone jack with a Lead II waveform at

0.2 V/mV of the ECG Lead II signal. The "HIGH LEVEL ECG" output is designed for

use with central station monitors or recorders requiring an analog high-level input.

Battery Test

Tests 8.4-volt and 9-volt batteries. The LED illuminates when the battery has more than a

6.2-volt charge.

Lead Test

Tests continuity of the ECG leads. The LED illuminates when the resistance is less than

1 ke.