Operation

RS-232 Serial Port Operation

3

3-63

Atrial Pacemaker Amplitude:

AHn

amplitude = n

n = 0, ±(2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 50, 100, 200, 500, and

700.)

Atrial Pacemaker Width:

AWn

width = n

n = 0.1, 0.2, 0.5, 1.0, and 2.0

Ventricular Pacemaker Amplitude:

VHn

amplitude = n

n = 0 ± (2, 4, 6, 8, 10, 12, 14, 16, 18, 20, 50, 100, 200, 500, and

700)

Ventricular Pacemaker Width:

VWn

width = n

n = 0.1, 0.2, 0.5, 1.0, and 2.0

Insert Premature Beat:

IPVC Insert

PVC

IPAC Insert

PAC

IPNC Insert

PNC

Respiration Waveform:

RNRM Normal

respiration

RVNT

Ventilator assisted respiration

Respiration Rate:

RRn

rate = n

n = 15, 20, 30, 40, 60, and 120

Respiration Ohms:

ROn

ohms = n

n = 0.0, 01, 0.2, 0.5, 1.0, and 3.0

Respiration Ratio:

RATn

ratio = n

n = 1/1, 2/1, 3/1, 4/1, and 5/1.

Respiration Apnea:

AOFF Apnea,

off

AON Apnea,

on

Al2

Apnea, 12 seconds

A32

Apnea, 32 seconds