Fluke Biomedical medSim 300B User Manual

Page 45

Advertising
background image

Operation

Operating the medSim 300B

3

3-17

premature arrhythmias

mf1

mf2

mf3

<F1>

<F2>

<F3>

<F4>

<

eic034.eps

The premature arrhythmia waveform selections can be made by pressing the

corresponding function key.
Descriptions of the premature arrhythmia selections follow:

Selections Descriptions
nsr

Normal sinus rhythm. Provides a convenient way to return to normal

sinus rhythm.

pac

Premature atrial contraction. Normal sinus rhythm except every 10

th

beat is premature by 25 %.

pnc

Premature nodal contraction. Nodal rhythm except every 10

th

beat is

premature by 25 %.

pvc

Premature ventricular contraction. Normal sinus rhythm except every

10

th

beat is a PVC.

mf1

Multifocal PVCs. Normal sinus rhythm except 2 of every 15 beats are

PVCs. PVC type alternates between type 1 and type 2.

mf2

Multifocal PVCs. Like "mf1" except PVC types are 2 and 3.

mf3

Multifocal PVCs. Like "mf1" except PVC types are 3 and 4.

The third group of arrhythmia waveforms is the ventricular rhythm 1 group—a group

with normal sinus rhythm at 80 BPM mixed with PVCs.
To select this group, press <F3>-VNT1 from the ARRHYTHMIAS MENU for the

following two display pages of six ventricular arrhythmia 1 waveforms:

ventricular arrh. 1

nsr

big

trig

pair

<F1>

<F2>

<F3>

<F4>

>

eic035.eps

ventricular arrh. 1

run5

run11

<F1>

<F2>

<F3>

<F4>

<

eic036.eps

Advertising