Power, Power-on conditions – Fluke Biomedical medSim 300B User Manual

Page 31

Advertising
background image

Operation

Power

3

3-3

Power

This is a battery-powered instrument with the option of using a battery eliminator.
The power control is a momentary "ON/OFF" key located on the top panel.
For optimum battery life, turn the instrument off when not in use, or use the battery

eliminator.
In addition, from the UTILITY MENU, you can preset a time interval to automatically

turn off the instrument after the last key has been pressed. The display flashes to indicate

a low battery condition when both batteries are low.
For battery replacement, the battery compartment is located on the bottom side of the

instrument. To access, twist the slotted latch one-quarter turn and lift. Fit a coin into the

slot for increased leverage. Replace both batteries at the same time with two 9-volt

alkaline batteries: Duracell MN1604 or equivalent.

Note

Do not use mercury, air, or carbon-zinc batteries.

Power-On Conditions

Upon power-on, the medSim 300B display shows the model number and the firmware

version.
The "power off time" (unless turned off) is displayed next. Following that, the Function

Select menu is displayed.
The following conditions are present when the power is first turned on:

Parameter Specification
ECG

Waveform: Normal sinus rhythm

Baseline BPM: 60

Axis:

Normal

Amplitude:1

mV

ST segment deviation: None

PVCs:

None

Pacemaker:Off
Respiration Waveform

Normal

Rate: 20 BrPM

Ohms:

1.0

e

Ratio:

3/1

Blood pressure

All channels: Atmosphere

Temperature 37

°C