Operators Manual

4-14

CATHETER

PA

RA

RV

Single lumen catheter in the pulmonary artery (PA)

eic172.eps

Figure 4-10. Pulmonary Artery Single Lumen Catheter Pullback

Note

The medSim 300B single lumen catheter pullback waveform for pulmonary

artery pressure simulates a measurement obtained with this procedure.

The following represents the simulated blood pressure waveforms for single and dual

lumen catheterization in the pulmonary artery:

Catheterization

Blood Pressure Waveforms

Single Lumen Catheter
Inserted

BP4 = RV Variable

Pulled back

BP4 = PA 25/10

Dual Lumen Catheter

Distal lumen (tip)

BP3 = PA 25/10

Proximal lumen

BP4 = RV Variable

To simulate a single lumen catheter with the lumen in the pulmonary artery:

1. Press <F3>-PULM from the CATHETERIZATION MENU, and then press

<F1>-PLBK.