medSim 300B

Operators Manual

1-8

either "positive" or "negative". The rate is shifted to

120 BrPM for 12 seconds each minute.

Ventilator Simulation:

40 BrPM at fixed ratio, other parameters variable as in

"normal"

Blood Pressure Section

Input/Output Impedance:

300 e

Exciter Input Voltage Range:

2 to 16 V

Exciter Input Frequency Range:

DC to 4 kHz

Output Sensitivity:

5 or 40

μ

V/V/mmHg

Output Range:

-10 to +300 mmHg on BP1, BP2 and BP4

-10 to +30 mmHg on BP3

Accuracy:

±

(1 % of full range + 1 mmHg) at 80 BPM, normal sinus

rhythm ECG only

Rate:

All dynamic pressures track all normal sinus rhythm rates

and track all arrhythmias

Isolation:

Blood pressure circuitry is electrically isolated from all

other medSim 300B outputs

Channel 1 (mmHg):

Atmosphere (0)

Arterial = 120/80

Left ventricle = 120/0

Central venous pressure = 15/10

Right ventricle = 25/0

Pulmonary artery = 25/10

Pulmonary artery wedge = 10/2

Static: -10, -5, 0, 20, 40, 80, 100, 200, 250, and 300

Manual or auto-stepping at 12 second intervals

Channel 2 (mmHg):

Atmosphere (0)

Arterial = 120/80

Left ventricle = 120/0

Central venous pressure = 15/10

Right ventricle = 25/0

Pulmonary artery = 25/10

Pulmonary artery wedge = 10/2

Static: -10, -5, 0, 20, 40, 80, 100, 200, 250 and 300

Manual or auto-stepping at 12 second intervals