Intra-Aortic Balloon Pump (IABP) Augmentation

During Intra-Aortic Balloon Pump (IABP) simulation, the medSim 300B generates an

arterial blood pressure waveform modified by the IABP "assist sync pulse".
You can send simulated normal sinus rhythms (NSRs), arrhythmias, or paced waveforms

to the intra-aortic balloon pump under test.
The medSim 300B also simulates various effects of balloon inflation and deflation timing

on arterial pressure, depending on how you have set the IABP controls.
The medSim 300B and the intra-aortic balloon pump under test are interactive during

IABP simulation.

Note

To produce medSim 300B simulations for IABP evaluation, you will need:

A blood pressure (BP) cable (refer to the current Fluke Biomedical

Price List for availability, part number information, and price).

A synchronization ("sync") cable (refer to the current Fluke

Biomedical Price List for availability, part number information,

and price).

ECG leads.

For IAPB simulation, connect the medSim 300B to the intra-aortic balloon pump under

test.
Follow the diagram below:

Arterial Blood
Pressure Input

IABP

medSim 300B

Assist Interval or
Sync Output

ECG Input

BP Cable

LV

Sync/A Pace
Input

Sync Cable

ECG Leads

BP4/AUX Output

eic204.eps

Figure 4-24. Connections for Intro-Aortic Balloon Pump (IABP) Simulation

Turn both units on.