Fluke Biomedical medSim 300B User Manual

Page 118

Advertising
background image

medSim 300B

Operators Manual

4-20

LUNGS

CATHETER

PA

LA

RA

RV

MITRAL VALVE

LAP is sensed
via the lungs.
(LAP PAW)

Catheter is wedged in the pulmonary capillary bed and
is blocking pressure from the right side of the heart.

LV

eic183.eps

Figure 4-15. Catheter wedged in the Pulmonary Artery (PA)

Note

The medSim 300B waveform for pulmonary artery wedge (PAW) simulates

a measurement obtained with this procedure.

To simulate a PAW pressure on the medSim 300B:

1. Press <F2>-MITR from the CATHETERIZATION MENU, and then press

<F1>-PAW3. The display will change to the one illustrated below:

bp1:lv

pa3

down

up

<F1>

<F2>

<F3>

<F4>

bp3:paw

10/2

eic184.eps

Advertising