Function select - ecg - pvc parameters menu – Fluke Biomedical medSim 300B User Manual

Page 49

Advertising
background image

Operation

Operating the medSim 300B

3

3-21

Descriptions of the PVC time selections follow:

Selections Descriptions
RonT

65 % premature (260 ms after R-wave).

early

33 % premature.

std

25 % premature (standard).

2. To set the number of PVCs per minute, press <F3>-/MIN from the PVC

PARAMETERS MENU:

pvc per minute: ##

up

down

exc

auto

<F1>

<F2>

<F3>

<F4>

eic044.eps

The number of PVCs per minute can be set from 0 to 25. Zero turns off the

function. To set, follow the instructions for the step menu functions.

3. To count PVCs, press <F4>-TOTL from the PVC PARAMETERS MENU:

PVC: ##,###

ET: ##:##.#

strt

off

hold

clr

<F1>

<F2>

<F3>

<F4>

eic045.eps

The TOTL MENU displays the number of PVCs generated and shows the

elapsed time (ET).

Descriptions of the "TOTL" executable selections follow:

Selections Descriptions
off

Turns off the counter, and clears the values.

strt

Initiates or restarts the counter.

hold

Stops the counter, but does not clear the values.

clr

Clears the values, but does not shut the counter off.