Operation

Operating the medSim 300B

3

3-31

bp 2 waveforms

rv

pa

paw

<F1>

<F2>

<F3>

<F4>

<

eic076.eps

2. Press the corresponding function key for the desired BP 2 dynamic waveform

selection. Waveform selections and their descriptions follow:

Selections Descriptions
atm

Atmosphere - 0 mmHg.

art

Arterial at 120/80 mmHg.

lv

Left ventricle at 120/0 mmHg.

cvp

Central venous pressure at 15/10 mmHg.

rv

Right ventricle at 25/0 mmHg.

pa

Pulmonary artery at 25/10 mmHg.

paw

Pulmonary artery wedge at 10/2 mmHg.

To select BP 2 static pressures, press <F2>-STAT from the BP 2 MENU for the

following display:

bp 2 static:

## mmHg

down

up

exc

auto

<F1>

<F2>

<F3>

<F4>

eic077.eps

Follow the instructions for the step menu functions presented earlier in this chapter.
To select BP 3, press <F3>-3 from the BP CHANNEL SELECT MENU to access the

following menu:

BP 3

WAVE

STAT

SWAN

<F1>

<F2>

<F3>

<F4>

eic078.eps