Fluke Biomedical medSim 300B User Manual
Page 59
Operation
Operating the medSim 300B
3
3-31
bp 2 waveforms
rv
pa
paw
<F1>
<F2>
<F3>
<F4>
<
eic076.eps
2. Press the corresponding function key for the desired BP 2 dynamic waveform
selection. Waveform selections and their descriptions follow:
Selections Descriptions
atm
Atmosphere - 0 mmHg.
art
Arterial at 120/80 mmHg.
lv
Left ventricle at 120/0 mmHg.
cvp
Central venous pressure at 15/10 mmHg.
rv
Right ventricle at 25/0 mmHg.
pa
Pulmonary artery at 25/10 mmHg.
paw
Pulmonary artery wedge at 10/2 mmHg.
To select BP 2 static pressures, press <F2>-STAT from the BP 2 MENU for the
following display:
bp 2 static:
## mmHg
down
up
exc
auto
<F1>
<F2>
<F3>
<F4>
eic077.eps
Follow the instructions for the step menu functions presented earlier in this chapter.
To select BP 3, press <F3>-3 from the BP CHANNEL SELECT MENU to access the
following menu:
BP 3
WAVE
STAT
SWAN
<F1>
<F2>
<F3>
<F4>
eic078.eps