medSim 300B

Operators Manual

3-28

2. Select the desired apnea mode by pressing the corresponding function key.

Apnea mode selections and their descriptions follow:

Selections Descriptions
off

Turns off apnea. Resumes breathing.

mom

Momentary. Causes apnea while the function key is pressed.

cont

Continuous apnea. Disabled by the "off" selection.

12 s

Generates 12 seconds of apnea each time the key is pressed.

32 s

Generates 32 seconds of apnea each time the key is pressed.

For respiration baseline shift:

1. Press <F2>-SHIFT from page 2 of the RESPIRATION MENU. Three selections

are available as shown:

resp base shift

off

pos

neg

<F1>

<F2>

<F3>

<F4>

eic067.eps

2. A selection of "positive" or "negative" shifts the respiration baseline either

positive or negative and changes the respiration rate to 120 BrPM for 12 of every

60 seconds.

Baseline shift selections and their descriptions follow:

Selections Descriptions
off

Turns off baseline shift.

pos

Shifts baseline up and decreases delta impedance to 1/6 for 12

seconds.

neg

Shifts baseline down and decreases delta impedance to 1/6 for 12

seconds.

Function Select - Blood Pressure Menu

To choose the BLOOD PRESSURE MENU, press <F3>-BP from page 1 of the

FUNCTION SELECT MENU.
The BP CHANNEL SELECT MENU offers the following five selections—1, 2, 3, 4 and

zero—on two pages of display:

BP CHANNEL SELECT

1

2

3

4

<F1>

<F2>

<F3>

<F4>

>

eic068.eps