4. At this menu level, the executable parameters for baseline BPM are selected. If,

for instance, a baseline BPM of 60 is desired, press <F2>-60, and the display

changes to:

step time (minutes)

0

0

0

.

save

<F1>

<F2>

<F3>

<F4>

*

*

eic116.eps

5. From the STEP TIME (MINUTES) MENU, set the desired run time for that

segment. Press the related function key repeatedly to increment the value directly

above that key.

6. Time is adjustable from 10 to 99 minutes using the <F1> and <F2> keys, and

from 0.0 to 9.9 minutes using the <F2> and <F3> keys.

Note

To cancel a setting, or to remove a selection, press the <up arrow> key, an

action that returns you to the previous menu. Then reselect the executable

parameter.

7. To save a desired setting, press <F4>-SAVE. When a time setting is saved, the

segment is placed into the sequence, and you are returned to the previous menu.

8. Continue in the same manner to program any additional segments into the

selected sequence. There can be up to 20 segments in a sequence.

9. After the last segment of a sequence is saved, use the <up arrow> key to return to

the FUNCTION SELECT MENU, page 2, then press <F3>-SEQ, and the

following menu appears:

save new sequence

stop

cycl

exit

<F1>

<F2>

<F3>

<F4>

*

*

eic117.eps

From this menu select the sequence ending:

Selections Descriptions
stop

Sequence holds at the last segment

cycl

Sequence cycles back to the beginning and continues

exit

Cancels the entire sequence just programmed and restores the previous

one

Note

An error beep results if an item that cannot be in a sequence (such as

viewing angle) is selected.