V pace, Sync/a pace, Bottom panel – Fluke Biomedical medSim 300B User Manual

Page 35: Battery compartment, Roms and cal switch access, Operating the medsim 300b, Initialization

Advertising
background image

Operation

Operating the medSim 300B

3

3-7

Note

The simulator can also be operated via the optional handheld controller.

Refer to theHHC3 Handheld Controller (PN2645641) manual for operating

instructions.

Menu Overview

Main Menu

The MAIN MENU—"FUNCTION SELECT"—offers eight main function groups, from

which all other selections are made.
The eight main function group menus are discussed in detail in the MENU

SELECTIONS section later in this chapter.
Located at the end of this chapter is a condensed listing of these menus.

Movement through Menus

When a menu choice is made by pressing the function key directly below the displayed

selection (<F1>, <F2>, <F3>, or <F4>), the display changes to show the menu for that

choice.
Using the function keys creates downward movement through the menus, one level at a

time. Pressing the <up arrow> moves back up through the menus, one level at a time.

Executable Parameter Values

At the lowest menu level are the executable parameter values. These executable

parameter values are indicated in lower case letters, such as, "baseline bpm - 30, 60, 80,

120, 160, 200, 240, 300"; while the other menus are presented in upper case letters: for

instance, "ECG - BASE, PERF, ARRH, PVC, PACE, INS".
A flashing display indicates that a selection is active. If an illegal key is pressed, three

quick beeps sound.
Some functions save the values in nonvolatile memory (EEPROM). Such a save is

indicated by a single long beep after a short delay.

Step Menus

Several functions have step menus. Step menus allow for stepping, either manually or

automatically, through the available parameter values.
Setting the parameter for the ECG baseline amplitude exemplifies how a step menu

functions. To illustrate—from the FUNCTION SELECT MENU:

1. Press <F1>-ECG, then press <F1>-BASE, and then press <F2>-AMP.

Your display now appears like this:

ecg amp, lead II: #.## mv

down

up

exc

auto

<F1>

<F2>

<F3>

<F4>

eic004.eps

Advertising