Operation
Operating the medSim 300B
3
3-11
Function Select - ECG - Baseline ECG Menu
To access the BASELINE ECG MENU:
1. Press <F1>-BASE from page 1 of the ECG MENU.
The five parameter groups for setting the normal sinus rhythm ECG are displayed
in two pages as shown in the following two illustrations:
BASELINE ECG
BPM
AMP
AXIS
ST
<F1>
<F2>
<F3>
<F4>
>
eic012.eps
BASELINE ECG
NEO
<F1>
<F2>
<F3>
<F4>
<
eic013.eps
2. From page 1 of the BASELINE ECG MENU, press <F1>-BPM. The eight
"baseline bpm" parameter values are shown on two pages as follows:
baseline bpm
30
60
80
120
<F1>
<F2>
<F3>
<F4>
>
eic014.eps
baseline bpm
160
200
240
300
<F1>
<F2>
<F3>
<F4>
<
eic015.eps
Selections made from this BASELINE BPM MENU set the normal sinus rhythm
rate (canceling arrhythmias and restoring all previously set baseline values).