Cardiac output rs-232 serial port operation – Fluke Biomedical medSim 300B User Manual

Page 145

Advertising
background image

Cardiac Output

Cardiac Output RS-232 Serial Port Operation

5

5-9

2. Press the corresponding function key for the desired C.O. trend sequence.

Selections of C.O. trend sequences and listings of their segments follow:

Selections Descriptions
NORM

5.0, 4.5, 6.0, 5.5 and 4.0 L/min

INTR

4.5, 5.5, "intr", 5.5 and 4.5 L/min

SLOW

4.5, 5.5, "slow", 5.5 and 4.5 L/min

Upon pressing a function key related to one of the three C.O. trend sequences—NORM,

INTR, or SLOW—the top line of the display shows the segments of that particular trend

sequence.
For illustration purposes:

1. Press <FI>-NORM and this appears:

5.0

strt

rset

<F1>

<F2>

<F3>

<F4>

4.5

6.0

5.5

4.0

eic223.eps

2. One of the segments (in this case, 5.0) has a cursor under it. Press <F1>-STRT to

execute the marked segment. Upon execution, the marked segment (5.0) starts

flashing, and the cursor marks the next segment (4.5).

3. Press <F1>-STRT again to execute the next marked segment (4.5). The cursor

then proceeds to mark the next segment (6.0). Continue in this manner to

complete the sequence.

Note

Flashing stops at the completion of a segment.

4. To reset the sequence, press <F2>-RSET and the cursor returns to mark the first

segment (5.0).
The three C.O. TRENDS MENU choices all follow the same pattern of menu

operation.

Cardiac Output RS-232 Serial Port Operation

By category, the following is a list of the medSim 300B Cardiac Output Option RS-232

remote commands and their descriptions.

Note

For complete information on remote commands, refer to the “RS-232 Serial

Port Operation” section in Chapter 3, and to the HHC3 Handheld

Controller Reference Manual..

Note

Each command is preceded by a bullet (•), which is not part of the

command string.