Operation

Operating the medSim 300B

3

3-15

To set R-wave amplitude or width parameters, follow step menu instructions as presented

earlier in this chapter.
Upon leaving R-wave detection, the ECG amplitude is equal to the adjusted R-wave

amplitude.

Function Select - ECG - Arrhythmias Menu

To select the ARRHYTHMIA MENU, press the <F3>-ARRH key from the ECG MENU:

ARRHYTHMIAS

SV

PREM

VNT1

VNT2

<F1>

<F2>

<F3>

<F4>

>

eic028.eps

ARRHYTHMIAS

COND

<F1>

<F2>

<F3>

<F4>

<

eic029.eps

Arrhythmias are divided into these five groups: supraventricular, premature, ventricular

rhythm 1, ventricular rhythm 2 and conduction.
Those that contain normal beats have an 80 BPM rate and unless otherwise stated, the PR

interval is 0.16 seconds.
When an arrhythmia contains one or more abnormal beats within a series of normal beats,

the abnormal beat is the second beat of the total series starting after one normal beat.
Supraventricular is a group of upper heart defect arrhythmias. To select the

supraventricular group, from the ARRHYTHMIAS MENU, press <F1>-SV for the

following three display pages of ten supraventricular waveform selections:

sv arrhythmias

nsr

af1

af2

aflt

<F1>

<F2>

<F3>

<F4>

>

eic030.eps

sv arrhythmias

sa

mb80

mb120

pat

<F1>

<F2>

<F3>

<F4>

>

<

eic031.eps