medSim 300B Instrument Specifications

1

1-5

ECG Performance Waveform Selections

Amplitudes are set by "Baseline ECG Amplitude". Amplitude applies to Lead II and

V-leads. Lead I is 0.7 times Lead II; Lead III is 0.3 times Lead II. The V-leads are the

same as Lead II.

Performance Waveforms

Rates

Zero Zero

output

Pulse 4

seconds

Square 2

Hz

Triangle 2

Hz

Sine

0.05, 0.5, 1, 10, 25, 30, 40, 50, 60, and 100 Hz

R-Wave Detection

Width: 20 ms increments from 20 to 200 ms and

two smaller widths of 8 and 12 ms. Auto-step time

interval is 6 seconds.

Amplitude: 0.05 mV to 0.5 mV in 0.05 mV steps;

0.5 mV to 5.5 mV in 0.25 mV steps. Auto-step

time interval is 6 seconds.

R-Wave Rate: 60 BPM

Arrhythmia Selections

Supraventricular

Atrial fibrillation (1 and 2)

Atrial

flutter

Sinus

arrhythmia

Missed beat at 80 BPM

Missed beat at 120 BPM

Paroxysmal atrial tachycardia (PAT)

Nodal

rhythm

Supraventricular tachycardia (SVT)

Premature

Premature atrial contraction (PAC)

Premature nodal contraction (PNC)

Premature ventricular contraction (PVC)

Multifocal

1

Multifocal

2

Multifocal

3

Ventricular Rhythm

Bigeminy

Trigeminy

Couplet (pair of PVCs)

Run of 5 PVCs

Run of 11 PVCs