Menu selections and their descriptions follow:

Selections Descriptions
start

Starts the series

hold

Disables the signal from the "SYNC/A PACE" input. When "hold" is

flashing, the signal is disabled

2. Press <F1>-START to start "Emergency I", and the display changes to:

emergency

1: running

start

hold

<F1>

<F2>

<F3>

<F4>

eic124.eps

When the waveform changes to ventricular fibrillation, the display changes to:

emergency

1: vfib

start

hold

defib

<F1>

<F2>

<F3>

<F4>

eic125.eps

3. Press <F3>-DEBB or input a pulse signal through the rear panel "SYNC/A

PACE" jack to defibrillate the simulator and convert the ECG to normal sinus

rhythm.

Note

For the medSim 300B patient simulator to be able to receive the input

signal, make certain that "hold" is not flashing.

A successful defibrillation is indicaled as the display changes to:

emergency

1: converted

start

hold

<F1>

<F2>

<F3>

<F4>

eic126.eps

The sequence can be repeated by pressing <F1>-START.