Function select - defibrillator tests menu – Fluke Biomedical medSim 300B User Manual
Page 80
medSim 300B
Operators Manual
3-52
Menu selections and their descriptions follow:
Selections Descriptions
start
Starts the series
hold
Disables the signal from the "SYNC/A PACE" input. When "hold" is
flashing, the signal is disabled
2. Press <F1>-START to start "Emergency I", and the display changes to:
emergency
1: running
start
hold
<F1>
<F2>
<F3>
<F4>
eic124.eps
When the waveform changes to ventricular fibrillation, the display changes to:
emergency
1: vfib
start
hold
defib
<F1>
<F2>
<F3>
<F4>
eic125.eps
3. Press <F3>-DEBB or input a pulse signal through the rear panel "SYNC/A
PACE" jack to defibrillate the simulator and convert the ECG to normal sinus
rhythm.
Note
For the medSim 300B patient simulator to be able to receive the input
signal, make certain that "hold" is not flashing.
A successful defibrillation is indicaled as the display changes to:
emergency
1: converted
start
hold
<F1>
<F2>
<F3>
<F4>
eic126.eps
The sequence can be repeated by pressing <F1>-START.