Fluke Biomedical medSim 300B User Manual

Page 67

Advertising
background image

Operation

Operating the medSim 300B

3

3-39

2. Press the corresponding function key for the desired lead selection.

Lead selections and their descriptions follow:

Selections Descriptions
all All
ra

Right arm

la Left

arm

11 Left

leg

vl Chest

To select ECG artifact size:

1. Press <F3>-SIZE from the ECG ARTIFACT MENU to access the three size

factors as shown in the following menu:

artif size (x LII ampl)

.25

.5

1

<F1>

<F2>

<F3>

<F4>

eic101.eps

The ECG artifact size factor multiplied by the ECG Lead II amplitude determines

the size of the ECG artifact. For example, 0.5 (artifact size factor) times 5 mV

(ECG Lead II amplitude) equals 2.5 mV (ECG artifact size).

2. To select the ECG artifact size, press the function key that corresponds to the

desired artifact size factor—0.25, 0.5, or 1.0.

Function Select – Artifact Blood Pressure/Respiration Artifact Menu

To access the BLOOD PRESSURE/RESPIRATION ARTIFACT MENU, press <F2>-

BP/R from the ARTIFACT MENU for this display:

bp/resp artif (% or mm)

off

5

10

<F1>

<F2>

<F3>

<F4>

eic102.eps

Respiration artifact can be injected into any blood pressure waveform.
Arterial and left ventricle waveforms are modulated by the selected respiration rate at

5 % or 10 %, while all others have a respiration artifact added to them at 5 mmHg or

10 mmHg.
Select 5 % or 5 mmHg by pressing <F2>-5, and select 10 % or 10 mmHg by pressing

<F3>-10.

Function Select - Sequences Menu

To access the SEQUENCES MENU, press <F3>-SEQ from page 2 of the FUNCTION

SELECT MENU for the following display: