Function select - sequences menu – Fluke Biomedical medSim 300B User Manual

Page 68

Advertising
background image

medSim 300B

Operators Manual

3-40

SEQUENCES

AUTO

STEP

<F1>

<F2>

<F3>

<F4>

eic103.eps

Sequences are preprogrammed lists of waveform selections that can be stepped through

either automatically or manually.
There are two types of sequences—autosequences and step-sequences.
Autosequences run automatically with no display interaction. There are eight different

selections in the AUTOSEQUENCE MENU.
Step-sequences can be started and stopped manually and involve screen display. In the

STEP-SEQUENCE MENU there are three preprogrammed step-sequences and four user-

programmable step-sequences.

Function Select - Sequences - Autosequences Menu

To select the AUTOSEQUENCES MENU:

1. Press <F1>-AUTO from the SEQUENCES MENU:

auto sequences

fail

vnt1

cond

resp

<F1>

<F2>

<F3>

<F4>

>

eic104.eps

auto sequences

vnt2

test

setup

start

<F1>

<F2>

<F3>

<F4>

<

eic105.eps

2. From these eight selections, choose an autosequence by pressing the

corresponding function key.

Autosequence selections are listed below along with their segments and run times. The

last entry in each list indicates whether the sequence stops, or returns to the first segment

and repeats. An autosequence can be terminated by pressing the <up arrow> key.

Selections Segments

Run

Times

fail (Cardiac failure)

ECG BPM 80

1.0 min

PVCs 12/min

0.2 min

PVCs

0/min