medSim 300B

Operators Manual

4-26

Illustrations of the periodic fetal heart rate selections, with their selection names and

descriptions follow:

Elapsed Time

0

90

100

140

mmHg

BPM

IUP Peak

Fetal Heart Rate

Intrauterine Pressure

eic194.eps

Figure 4-19. medSim 300B Simulated Early Deceleration

edec: Early deceleration. Fetal heart rate starts at 140 BPM, slows to 100 BPM at

intrauterine pressure peak, and returns to 140 BPM as the intrauterine pressure

falls back to zero.

Elapsed Time

0

90

100

140

mmHg

BPM

IUP Peak

Fetal Heart Rate

30 second lag

Intrauterine Pressure

eic195.eps

Figure 4-20. medSim 300B Simulated Uniform Deceleration

udec: Uniform deceleration. Fetal heart rate change lags intrauterine pressure change

by 30 seconds. Fetal heart rate starts at 140 BPM, slows to 100 BPM, and returns

to 140 BPM.