medSim 300B

Operators Manual

5-4

Trends: Normal

sequence

Sequence with interrupted injectate

Sequence with slow injectate

Accuracy:

10 % of setting for 3 L/min, 5 L/min, and 7 L/min at 37 °C

±0.4 °C for baseline temperature

Isolation:

Cardiac output is electrically isolated from the rest of the

unit except temperature channel 1

Note

Cardiac output is inoperative during the temperature channel 1 simulation

("VAR TEMP" mode).

Accessories

Standard

Accessories Part

No.

CI-3 module/cable

2392199

General purpose connector

2392158

Note

Additional general purpose connectors are available from Fluke

Biomedical.

Optional

Currently offered cardiac output adapter cables are listed below:

Hewlett Packard Cardiac Output Adapter

Hewlett Packard Adapter

Gould/SpectraMed Model 1445 Adapter

Gould/SpectraMed Model 1465 Adapter

Note

Refer to the current Fluke Biomedical Price List for availability, part

number information, and price.

Cardiac Output Simulation

Introduction

The Cardiac Output Option for the medSim 300B consists of both internal firmware and

external hardware.
The external hardware is designated as the CI-3 module/cable, and plugs directly into the

medSim 300B top panel at the "CARDIAC OUTPUT" jack.

Note

See illustration "CI-3 module/cable" located at the end of this chapter.