Fluke Biomedical medSim 300B User Manual

Page 115

Advertising
background image

Option 1

Cardiac Catheterization

4

4-17

Follow the illustration below.

Blood Pressure
Input

Cardiac
Catheterization
System

medSim 300B

Blood Pressure
Input

BP Cables

LV

BP3

PA OR PAW

Use BP1 and BP3 Blood Pressure
Channels for Mitral Valve Simulation

BP1

eic177.eps

Figure 4-13. Cardiac Catheterization Connections for Mitral Valve Simulation

1. Turn both units on.
2. After a connection has been made, the medSim 300B and the monitor need to be

zeroed.

3. To zero all four of the medSim 300B blood pressure channels, press <F3>-BP

from page 1 of the FUNCTION SELECT MENU, for the following two pages of

display:

BP CHANNEL SELECT

1

2

<F1>

<F2>

<F3>

<F4>

3

4

>

eic178.eps

BP CHANNEL SELECT

<

zero

CATH

<F1>

<F2>

<F3>

<F4>

eic179.eps

4. Press <F1>-ZERO from the second page of the menu.

This sets all the medSim 300B blood pressure channels to zero.

Note

Necessary to complete only once after connecting the medSim 300B to the

equipment under test (EUT).