Fluke Biomedical medSim 300B User Manual

Page 78

Advertising
background image

medSim 300B

Operators Manual

3-50

Function Select - Sequences -Step-sequences - User Sequences Menu

After programming user step-sequences, from the USER SEQUENCES MENU choose

the user step-sequence to run.
To select the USER SEQUENCES MENU:

1. Press <F2>-USER from the STEP SEQUENCES MENU for the following menu

display:

USER SEQUENCES

SEQ1

SEQ2

SEQ3

SEQ4

<F1>

<F2>

<F3>

<F4>

eic120.eps

2. Press the corresponding function key to select the desired sequence.

Note

If a selected sequence has not yet been programmed, an error beep sounds.

To program user step-sequences, follow the instructions that appear in the

Make Sequences Menu section located earlier in this chapter.

Upon pressing a function key related to a previously programmed user step-sequence, the

display, as represented below, provides the following information:

S# ##:##.#

????

##.#

strt

stop

next

rset

<F1>

<F2>

<F3>

<F4>

eic121.eps

From the top line of the display:

S#

Is the sequence name: "S1", "S2", "S3", or "S4"

##:##.#

Is the sequence run time

????

Is the segment name (abbreviated)

##.#

Is the segment run time

Menu operations for user step-sequences are the same as for preset (or fixed)

step-sequences.
The user step-sequence selections and descriptions follow:

Selections Descriptions
strt

Starts the sequence running

stop

Stops the sequence

next

Steps to the next segment if running

rset

Reset stops the sequence, returns to the first segment, and clears the

time. Reset does not restart the sequence