Panel descriptions, controls and connectors, Top panel, Lcd display and keyboard – Fluke Biomedical medSim 300B User Manual

Page 33: Ecg leads, Rs232, High level ecg, Battery test, Lead test

Advertising
background image

Operation

Panel Descriptions, Controls and Connectors

3

3-5

Cardiac Output

"CARDIAC OUTPUT" is optional. When installed, refer to the "Cardiac Output" chapter

for operating instructions.

Temperature—Variable and Fixed

For variable temperature selection, set the "CARDIAC OUTPUT/ VAR TEMP" switch to

"VAR TEMP"

*

, and adjust the temperature from the keyboard. Set the "PROBE" switch

to match the monitor probe type-400 or 700

**

.

Note

*

"CARDIAC OUTPUT' is disabled in "VAR TEMP" mode.

**

Temperature connector cables are available from Fluke Biomedical. The

type of probe simulated—either the 400 or 700 series YSI—determines the

type of cable to use.

Fixed temperature ("FIXED TEMP") is 37 °C/98.6 °F.

Blood Pressure

There are four blood pressure channels, labeled as "BP1", "BP2", "BP3" and

"BP4/AUX". Each channel is independently adjustable from the keyboard.
Set the "SENSITIVITY" switch to match the patient monitor's input sensitivity (either 40

or 5 µV/V/mmHg).

Note

To connect to the four blood pressure connectors, cables prewired for many

monitors, as well as an unterminated cable (see illustration "Unterminated

Blood Pressure Output Cable Assembly" at the end of this chapter), are

available from Fluke Biomedical. Refer to the medSim 300B Accessories

section in Chapter 1 and to the Fluke Biomedical Price List.

Respiration

The respiration signal is output across the ECG leads.

Corresponding with the type of patient monitor in use, set the "LEAD SELECT"

switch to the lead in use (I or II/RL-LL).

Use the "BASELINE IMPEDANCE" switch to set the impedance (500, 1000,

1500, or 2000) between any two limb leads.

Use the keyboard to set the respiration parameters.

Rear Panel

Battery Eliminator

The jack ("BATT ELIM 9 V DC") for the battery eliminator on the medSim 300B rear

panel. Use only a battery eliminator supplied by Fluke Biomedical.

Note

Refer to the medSim 300B Accessories section in Chapter 1 and to the

Fluke Biomedical Price List.