5 pump does not start, 6 no flow, Pump does not start – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 102: No flow, Section 4.6, For pump

Advertising
background image

ICS-2100 Ion Chromatography System

90

Doc. 065291-01 3/09

4.5

Pump Does Not Start

No power (front

Power

LED indicator is not lit)

Check that the power cord is plugged in.

Check the main power fuses and replace, if needed (see

Section 5.18

).

No communication between ICS-2100 and Chromeleon

Verify that the USB cable is connected correctly. For connection and setup
information, refer to the ICS-2100 installation instructions.

Pump is turned off

Turn on the pump from either the Chromeleon Control panel or the ICS-2100
touch screen

HOME

page.

Flow rate is set to 0

Set the flow rate from either the Chromeleon Control panel or the ICS-2100
touch screen

HOME

page.

4.6

No Flow

Pump waste or priming valve open (see

Figure 4-2

)

Close the valves by turning the knobs clockwise until fingertight. DO NOT
OVERTIGHTEN! Overtightening may damage the valve and the pump
head.