A.15 autosampler, A.16 system software, A.15 autosampler a.16 system software – Oritron Dionex Ion Chromatography System ICS-2100 User Manual

Page 216: Ics-2100 ion chromatography system

Advertising
background image

ICS-2100 Ion Chromatography System

204

Doc. 065291-01 3/09

A.15 Autosampler

A.16 System Software

Void Volume

SRS 300 (4 mm): <50

μL

SRS 300 (2 mm): <15

μL

MMS 300 (4 mm): <50

μL

MMS 300 (2 mm): <15

μL

AMMS ICE 300 (4 mm): <50

μL

AMMS ICE 300 (2 mm): <15

μL

Anion AES: <35

μL

Cation AES: <35

μL

Automation

Using

Autosampler

Dionex AS, AS-DV, or AS-HV Autosampler (or third-party
autosampler)

Sequential/

Simultaneous

Injection

May be available; depends on autosampler capabilities

Automated

Dilution

Available with AS Autosampler

Dilution Factor: 1:1 to 1:1000

Dilution Time: 15 seconds with sample overlap

Online Sample

Degassing

Optional; CRD 200/300 required

Online Filtration

Available with an AS-DV Autosampler or an inline filter

Automation

Flexibility

High, using Chromeleon software and post-run features

Software

Chromeleon Chromatography Management System or Chromeleon
Xpress; validated for use with Windows Vista or Windows XP

Automated

Procedure

Wizards

Standard feature

System Smart

Startup and

Shutdown

Standard feature